The technical alarm condition (outside the rated range) is preset
in the factory and cannot be adjusted or inactivated. This alarm
condition is assigned as low priority according to IEC 60601-1-8.
The technical alarm is non-latching and need no reset. The
signal displayed on LCD will disappear automatically after about
8 seconds.
11. Troubleshooting (1)
Problem
Possible cause
The cuff position was
not correct or it was not
properly tightened.
Body posture was not
correct during testing.
LCD Display
shows
Speaking, arm or body
abnormal
movement, angry,
result
excited or nervous
during testing.
Irregular heartbeat
(arrhythmia)
160
PC-BMG3018_IM
Solution
Apply the cuff correctly
and try again.
Review the section
"Body posture during
measurement" of the
instructions and try
again.
Re-test when calm and
without speaking or
moving during the test.
It is inappropriate for
people with serious
arrhythmia to use this
electronic sphygmo-
manometer.
04.07.18