Indications For Use; Technical Data - Puritan Bennett Breeze SleepGear Guide D'utilisation

Masquer les pouces Voir aussi pour Breeze SleepGear:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 15

Indications for Use

This product is intended for use with devices that deliver continuous
positive airway pressure (CPAP) and bi-level positive airway
pressure in patients weighing at least 30 kg. Refer to the warnings
on page iv.

Technical Data

The following table provides the technical data for both the
Nasal Pillows and DreamSeal configurations.
Vent Flow, Pressure Range, and Materials
CPAP Pressure
Parameter
Pressure Range
Vent Flow
Storage and
Operating Conditions
Relative Humidity
(noncondensing)
Materials
Breeze SleepGear
(cmH
O)
2
3 – 30
with DreamSeal
3
13
5
16
8
21
10
23
12
25
15
28
18
31
20
32
25
34
30
38
Storage
Temp. Range
-40°F to 158°F
(-40°C to 70°C)
Up to 95%
• Nasal Pillows: medical-grade silicone
• DreamSeal Perimeter Seal: medical-grade silicone
• Halo Straps: Velstretch
®
Overview
E n g l i s h
Flow Rate
(L/min)
with Nasal Pillows
17
20
25
28
31
35
39
41
47
53
Operating
41°F to 104°F
(5°C to 40°C)
Up to 95%
3

Publicité

Table des Matières
loading

Table des Matières