Indications for Use

Technical Data

Breeze SleepGear
This product is intended for use with devices that deliver continuous
positive airway pressure (CPAP) and bi-level positive airway
pressure in patients weighing at least 30 kg. Refer to the warnings
on page 4.
The following table provides the technical data for both the Nasal
Pillow and DreamSeal configurations.
Vent Flow, Pressure Range, and Materials
Parameter
Vent Flow
Pressure Range
Materials
CPAP Press.
(cm water)
3
5
8
10
12
15
18
20
25
30
35
3 – 35 cmH
O
2
Nasal Pillows: medical-grade silicone
DreamSeal Perimeter Seal: medical-
grade silicone
Halo Straps: Velstretch
Overview
E n g l i s h
Flow Rate
(L/min)
13
16
21
23
25
28
31
32
34
38
40
®
7