1. Applications
The SE 603 conductivity sensor is characterized by maximum corrosion resis-
tance for high-pressure and high-temperature measurements. The sensor has a
universal measuring range from ultrapure water to 600 mS/cm.
Process-wetted materials: platinum and PTFE. A temperature detector is
integrated in the sensor. Electrical connection is made via fixed cable.
Fields of application: Highly corrosive chemical processes, bleaching liquors,
oxidizing as well as heavily polluted (e.g. fibrous) materials, condenser leakage
monitoring.
2. Dimension Drawing
23
Cable length
approx. 5 m
3. Calibration / Adjustment
Remove sensor, clean, rinse, and dry it, then immerse it in a calibration solution.
Suitable calibration solution: ZU 0348 (12.88 mS/cm). Perform a product
calibration if the sensor can or shall not be removed. Observe instruction
manual of meter. The cell constant of the sensor can be certified by the manu-
facturer (ZU 0320). For that purpose, you must send in the sensor.
10
14
61
53
2.85
20°
Assembly
ZU 0218 and
ZU 0219
are required
165
:
Field area
34