Chauvin Arnoux C.A 6549 Mode D'emploi page 72

Masquer les pouces Voir aussi pour C.A 6549:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1
§
Default Probe Temperature
§
Rc Reference Temperature
∆T for R/2
§
Calculate ∆T fr
§
T fr
T fr
T fr
T from Memor
om Memor
om Memor
om Memory y y y y
om Memor
§
Maximum Output Voltage
§
Set Default Parameter
§
Clear Memory
§
V Disturbance / V Output = dlSt factor (cf. § 3.2 - «important remark» )
§
Buzzer
§
Power Down
estimated measurement temperature unit
Default value
23°C
reference temperature to which the measurement result must be referred
Default value
40°C
estimated ∆T to obtain an insulation resistance / 2
Default value
10°C
used to calculate ∆T from 3 stored measurements performed on the same
instrument but at different temperatures (see § 4.5.3)
imposes maximum test voltage
Default value
5000V
default configuration : reinitializes the instrument with the default values
of all parameters.
can be used to partially or completely erase stored data cf. § 4.5.2
Default value
3%
activation / desactivation of buzzer (keys, measurements, alarms)
Default value
ON
automatic shut-down of the instrument after 1mn
if no key is activated
Default value
OFF
72
Range
-15°C... +75°C
Range
-15°C... +75°C
Range
-15°C... +75°C
Range
40... 5100V
Range
3%, 10% or 20%
Range
ON or OFF
Range
ON or OFF

Publicité

Table des Matières
loading

Table des Matières