Accessories; Product Contact Parts - IKA VC 10 Mode D'emploi

Masquer les pouces Voir aussi pour VC 10:
Table des Matières

Publicité

Description
Warning
Temperature Calibration
Error
Warning
Pressure Calibration Error
Error 3
Device Temperature Error.
Error 9
Logic Storage Error.
Display Storage Error.
Error 68
Pressure Out Of Range.
Error 69
Pressure Sensor Error.
Error 71
Internal Communication
Error
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
contact the service department,
send the instrument for repair, including a short description of the fault.
Vacuum pump IKA
MVP 10 basic
®
Vacuum pump IKA
MVP 10 digital
®
Temperature sensor IKA
Pump control IKA
®
VC 2.4
Set of hoses IKA
RV 10.5005
®
Analogue cable MVP 10.100
Designation
Connection spigots
Distributor
Pressure sensor
Vacuum valve
Venting valve
Detection
Temperature calibration value is out
of range.
Pressure calibration value is out of
range.
Analysis of internal PCB temp.
Sensor.
Inside temperature reaches limit
value.
Read or write internal memory
error.
Pressure is out of range.
Pressure is bigger than room
pressure.
Pressure sensor output is too low.
Internal Watchdog elapsed.
PT 1000
®

Product contact parts

Material
PP
PPS
FPM/AL2O3
PEEK; EPDM; FKM
PEEK; EPDM; FKM
Reason
Wrong simulator resistance is chosen.
Set wrong temperature calibration
value.
Input wrong vacuum actual value
when calibration.
Set wrong vacuum calibration value.
Sensor is broken.
Room temperature > 40 °C.
PCB or valve is broken.
Read or write internal memory error.
Pressure in the recipient is too high.
Sensor is broken.
Sensor is not connected.
Sensor is broken.
Communication lost between logic
board and display board.

Accessories

VC 10.300 Check valve
VC 10.400 Stand
VH.SI.8 Vakuum hose
for further accessories see www.ika.com.
Action
Check the simulator resistance.
Call service department.
Calibrate again.
Check the set calibration value.
Call service department.
Switch off device for cool down.
Call service department.
Call service department.
Check the air flow of pump and vent
recipient.
Call service department.
Check the sensor connection.
Call service department.
Restart device (unplug mains cable,
plug in mains cable).
Check communication between logic
board and display board.
Call service department.
49

Publicité

Table des Matières
loading

Table des Matières