Device Serviceability Check (Testing) - RIKTA ESMIL Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

RIKTA
– ESMIL
(2A)
®
®

DEVICE SERVICEABILITY CHECK (TESTING)

If the Device had been off for more than 8 hours then when the Device is turned
on, the serviceability check standby mode will be activated automatically. At
that all frequency [1] indicators and all time [4] indicators will light up.
and
buttons would not respond to pressing.
For the Device serviceability check:
Place the device so that the device aperture [10] is not directed to
a light source.
If the indicators «5 Hz» and «2
min» only light up after the beep,
device is functional and ready
for use (see «Conducting the
therapy session», page 36).
Briefly press and release
Pulse frequency indicators [1] and all treat-
ment time indicators [4] will turn off in turn.
The Device testing process is in progress.
Conducting
the therapy
session
page 36
If all pulse frequency LED indicators [1]
and all treatment time LED indicators [4]
blink and you hear an intermittent sound
signal, it means the Device is faulty.
Turn it off immediately and contact the
manufacturer or service center for as-
sistance.
button.
The Device
is faulty
35
EN

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Esmil 2a

Table des Matières