RIKTA
– ESMIL
(2A)
®
®
WHEN USING THE DEVICE
■
The Device falls within Class 1M laser equipment.
When using the Device, it is recommended to avoid eyes exposure
to direct or reflected radiation. Do not stare into the beam and do
not make direct observations using optical instruments.
■
Medical personnel are recommended to wear goggles while operating
the Device (available on request). It is allowed not to use goggles when
carrying out a single treatment procedure on condition that the Device
aperture is tightly pressed to a patient's body.
If the treatment is carried out in a patient's face area with the help of
light guide nozzles, the goggles are also recommended for the patient.
■
Start the laser therapy procedure only after putting the Device aper-
ture on the treatment zone.
DEVICE CARE
■
Before and after a treatment procedure, make sure to wipe the Device
with a cloth of gauze or coarse calico soaked in a treatment tool disin-
fectant based on 3% hydrogen peroxide solution. Avoid Device contact
with alcohol-containing media!
■
The Device disinfection may be carried out only when the Device is
turned off and disconnected from mains.
OPERATION CONDITIONS
The Device is designed for operation at a temperature from
+35°С
+10 to +35˚C, atmospheric pressure from 630 to 800 mm Hg,
+10°С
and relative humidity not more than 80%.
MEDICAL CONTRAINDICATIONS TO LASER THERAPY
It is STRICTLY FORBIDDEN to carry out laser therapy at home in
presence of the following medical contraindications:
– individual intolerance of influence factors of the device;
– general serious condition of the patient;
– blood diseases with the syndrome of bleeding;
– thrombocytopenia below 60,000 / µl;
29
EN