DEVICE SERVICEABILITY CHECK (TESTING)

If the Device had been off for more than 45 minutes then when the Device is
turned on, the serviceability check standby mode will be activated automati-
cally. At that all frequency [2] indicators and all time [4] indicators will light up.
FREQUENСY and TIME buttons would not respond to pressing.
Make sure that all frequency [2] and time [4] indicators
If the indicators blink the serviceability check can not be carried out. As a rule,
it happens if the emitting window [9] of the Device is targeted at a light source.
Position the Device so that the emitting window [9] is not targeted at a light
source. When the indicators stop blinking it is possible to proceed to the Device
serviceability check.
For the Device serviceability check:
If after the audible signal only
"64 Hz" and "5 min" indicators
are on the Device is service-
able and ready for operation
(see "Holding Treatment Ses-
sion", page 32).
are on but do not blink!
Briefly press and depress START/STOP button.
The Device testing process is in progress.
Frequency [2] and time [4] indicators will
go out one at a time.
Holding
treatment
sessions
page 32
If all frequency [2] and time [4] indicators
blink and intermittent audible signal is
heard it means that the Device faulty.
Immediately turn off the Device and resort
to the manufacturer or the service center.
intermittent
audible signal
The Device
is faulty