(11)
(3)
(7)
Camille Bauer AG
CH-561 0 Wohlen
Switzer land
Span
Zero
ON
(13) (6)
(12)
Fig. 2
(3)
Front plate
(6)
Type label
(7)
Transparent cover
(8)
Fixing bracket
(9)
Opening for withdrawing clip (for opening the housing)
(10)
Top-hat rail 35 × 15 mm or 35 × 7.5 mm (EN 50 022)
(11)
Terminals
(12)
Terminals
(13)
Space for notes
ON
Green LED for indicating device standing by
6. Technical data
Measuring input
DC current:
Standard ranges
0...20 mA, 4...20 mA, ± 20 mA
Limit values
0...0.1 to 0...50 mA
also live-zero,
start value > 0 to ≤ 50% fi nal
value
– 0.1...0...+ 0.1 to
– 50...0...+ 50 mA
also bipolar asymmetrical
R
DC voltage:
Standard ranges
0...10 V, 2...10 V, ± 10 V
Limit values
0...0.06 to 0...40, Ex max. 30 V
also live-zero,
start value > 0 to ≤ 50% fi nal
value
– 0.06...0...+ 0.06 to
– 40...0...+ 40 V,
Ex max. – 30...0...+ 30 V
R
16
(10)
Span
Zero
without
function
ON
= 15 Ω
i
= 100 kΩ
i
Overload:
(8)
(9)
Measuring output
DC current:
Burden voltage:
External resistance:
DC voltage:
Burden:
Current limiter at
R
max.:
ext
Voltage limiter at
R
= ∞:
ext
Residual ripple in
output current:
Response time:
Power supply H
AC/DC power pack (DC and 45...400 Hz)
Table 1: Nominal voltages and tolerances
Nominal voltage
U
N
24... 60 V
DC / AC
85...230 V
1
DC / AC
24... 60 V
DC / AC
85...230 V
AC
85...110 V
DC
1
For power supples > 125 V, the auxiliary circuit should include an
external fuse with a rating ≤ 20 A DC.
Power input:
Accuracy data (acc. to DIN/IEC 770)
Basic accuracy:
DC current
continuously 2-fold
DC voltage
continuously 2-fold
Standard ranges
0...20 mA, 4...20 mA, ± 20 mA
Limit values
0...1 to 0...20 mA
0.2...1 to 4...20 mA
– 1...0...+ 1 to – 20...0...+ 20 mA
12 V
R
max. [kΩ] =
I
ext
AN
I
= Output circuit full-scale value
AN
Standard ranges
0...10 V, 2...10 V, ± 10 V
Limit values
0...1 to 0...10 V
0.2...1 to 2...10 V
– 1...0...+ 1 to – 10...0...+ 10 V
≥ 2 kΩ
Approx. 1.1 ×I
for current output
AN
Approx. 13 V
< 0.5% p.p.
< 50 ms
Instrument
Tolerance
DC – 15...+ 33%
Standard
AC ± 15%
(Non-Ex)
DC – 15...+ 33%
AC ± 15%
Type of protection
± 10%
"Intrinsically safe"
[EEx ia] IIC
– 15...+ 10%
≤ 1.2 W resp. ≤ 3 VA
Limit error ≤ ± 0.2%
Including linearity and reproduci-
bility errors
12 V
[mA]
version