IEC60601-1-2
Immunity test
test level
3 Vrms
150 kHz to
Conducted RF
80 MHz
IEC 61000-4-6
80% AM (2Hz)
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
80% AM (2Hz)
Note1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects, and people.
*Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field
strength in the location in which the HBP-1100 is used exceeds the applicable
RF compliance level above, the HBP-1100 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the HBP-1100.
**Over the frequency range 150 kHz to 80MHz, field strengths should be less than 3 V/m.
Compliance
Electromagnetic environment -
level
guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the HBP-1100, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommend separation distance
d = 1,2 P
d = 1,2 P
d = 2,3 P
3 Vrms
where P is the maximum output
power rating of the transmitter
in watts (W) according to he
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
3 V/m
Field strengths from fixed RF
transmitters as determined
by an electromagnetic site
survey*, should be less than
the compliance level in each
frequency range**.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
150 kHz - 80 MHz
80 MHz - 800 MHz
800 MHz - 2,5 GHz
EN
37