Electromagnetic Immunity (IEC60601-1-2)
IEC60601-1-2
Immunity test
test level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
±2 kV for
Electric fast
power supply
transient/burst
lines
IEC 61000-4-4
±1 kV for
input/output lines
±1 kV
Surge
differential mode
IEC 61000-4-5
±2 kV
common mode
<5 % U
Voltage dips, short
for 0.5 cycle
interruptions and
voltage variations
40 % U
on power supply
for 5 cycles
input lines
70 % U
IEC 61000-4-11
for 25 cycles
U
: Rated voltage
<5 % U
T
of test unit
for 5 sec.
Power frequency
(50/ 60 Hz)
3 A/m (r.m.s)
magnetic field
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
T
36
Compliance
level
±6 kV contact
±8 kV air
±2 kV for
power supply
lines
±1 kV for
input/output lines
±1 kV
differential mode
±2 kV
common mode
<5 % U
T
T
for 0.5 cycle
40 % U
T
T
for 5 cycles
70 % U
T
T
for 25 cycles
<5 % U
T
T
for 5 sec.
3 A/m (r.m.s)
Electromagnetic envi-
ronment - guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30 %.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
HBP-1100 requires
continued operation
during power mains
interruptions, it is
recommended that the
HBP-1100 be powered
from an uninterruptible
power supply or
batteries.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.