Note that the new input value are put in effect only at the next
STOP/START of the pump.
If PRINT is selected and a printer is connected on RS 232 line, an
unidirectional communication is established and every minute the
data are sent to the printer, even if the pump is not running.
The set of data available are:
The controller is factory-set to HOST.
Turbo-V 700HT Controller User Manual / 87-900-877-01 (F)
Life time and cycle # zeroing
Configuration parameter readings
Configuration parameter setting
Pump speed KRPM
Pump temperatures
Pump current A
Pump power W
R1 condition
R2 condition
Confirm the selection by pressing the PUMP CURRENT push-
button.
In this way you enter into an operating phase named "Monitor
Relay Programming" described in the following paragraph.
Technical Information
Use
281/304
17