Varian PTB Turbo V 700 HT 969 9445 Notice D'emploi page 235

Masquer les pouces Voir aussi pour PTB Turbo V 700 HT 969 9445:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 41
MESSAGE
FAULT: SHORT
CIRCUIT
SYSTEM OVERRIDE
OVERVOLTAGE
Turbo-V 700HT Controller User Manual / 87-900-877-01 (F)
DESCRIPTION
After the starting phase the
output connection is shorted
(output current higher than 16
A).
The pump is stopped by an
emergency stop signal provided
via a remote contact.
Controller power supply circuitry
is faulty, or the Controller
received a spike.
Instructions for Use
Error Messages
REPAIR ACTION
Check connections and
shortages between pump
and controller. Press the
START push-button twice
to start the pump.
Remove the controller
power cable and check the
emergency condition. Then
reconnect the power cable
and press the START push-
button twice to start the
pump.
Press the START push-
button twice to start the
pump.
Should the message still be
present, call the Agilent
service
235/304