CanalPro
CL2 Cordless Handpiece
TM
6.3.4 Remove the old batteries and replace with the new batteries using care to
6.3.5 Re-install the cover (using hardware previously removed). Avoid over
7.0

Troubleshooting

Problem
LED Does not turn
on.
Battery not properly
charging
Charging base is not
working
Handpiece
overheating on
charging base
Handpiece head not
rotating
Handpiece power
weak
Auto/Reverse
function not working
RPM of handpiece
slow
16
observe the correct polarity.
tightening the screw. Place the unit on the charger until it indicates that
the battery is fully charged.
Cause
1. Battery completely drained
2. Handpiece not used for an
extended period of time.
1. Battery completely drained
2. Battery temp too low.
3. Battery temp too high.
4. Battery voltage is too high.
5. Improper contact with
charging base.
6. Error code is displayed
1. Power cord not plugged in.
2. Power is switched Off.
3. Blown Fuse.
Battery or charging base is
defective.
Debris in motor head
Battery is low
Battery is low
Battery is low
Solution
If the battery is not completely
drained, it must be recharged. If it
is completely drained it must be
replaced.
1. Replace the battery
2. Place the charger in an area
with proper temperature
control above 0° C (32° F).
3. If the battery temperature is
too high and the battery will
not charge, contact customer
service.
4. Ensure the correct batteries are
being used.
5. Ensure that the handpiece
is properly mounted to the
charging base.
6. Check the error message.
1. Check power connections
2. Turn power switch on.
3. Contact customer service.
Contact your local dealer.
Clean or replace the contra angle.
Charge the battery
Charge the battery
Charge the battery