CanalPro
CL2 Cordless Handpiece
TM
Handpiece
Charging LED
Note the following:
4.3
The CanalPro
10
•
If the charging LED is not lit after the power is turned on, there may
be a problem with the base circuit. Contact Customer Service for
instructions.
•
If the charging base terminals are dirty and contaminated, the handpiece
may not properly charged. Ensure all connections are clean prior to
using the charging base.
•
Use only the charging base supplied with the CanalPro
•
While the battery is charging the rear of the handpiece may become
warm which will not generally cause a problem.
•
Once the handpiece is placed on the charging base it is recommended
that it stays in place until the battery indicates fully charged.
•
A completely drained battery may not be able to be charged and may
have to be replaced.
•
Avoid placing the charging base in an area that may be susceptible to
additional heat (in the sunlight or next to a heater, etc.).
•
The battery may not charge if the temperature is excessively high or
low, it is already fully charged or if the battery is damaged.
CL2 Modes
TM
•
Auto Stop Reverse Forward
overloaded it will begin to run in reverse until the overload is relieved.
Power Switch
TM
– In this mode if the handpiece is
CL2.