Télécharger Imprimer la page

The Canalpro Tm Cl2 Modes - Coltene CanalPro CL2 Notice D'utilisation

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 56
CanalPro
CL2 Cordless Handpiece
TM
Handpiece
Charging LED
Note the following:
4.3
The CanalPro
10
If the charging LED is not lit after the power is turned on, there may
be a problem with the base circuit. Contact Customer Service for
instructions.
If the charging base terminals are dirty and contaminated, the handpiece
may not properly charged. Ensure all connections are clean prior to
using the charging base.
Use only the charging base supplied with the CanalPro
While the battery is charging the rear of the handpiece may become
warm which will not generally cause a problem.
Once the handpiece is placed on the charging base it is recommended
that it stays in place until the battery indicates fully charged.
A completely drained battery may not be able to be charged and may
have to be replaced.
Avoid placing the charging base in an area that may be susceptible to
additional heat (in the sunlight or next to a heater, etc.).
The battery may not charge if the temperature is excessively high or
low, it is already fully charged or if the battery is damaged.
CL2 Modes
TM
Auto Stop Reverse Forward
overloaded it will begin to run in reverse until the overload is relieved.
Power Switch
TM
– In this mode if the handpiece is
CL2.

Publicité

loading