Cleaning And Storing The Device And Cuff; Technical Specifications - Beurer BM 47 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 33
• the cuff is fastened too tightly or loosely (
appears on the display)
• the pump pressure is higher than 300 mmHg (
appears on the display)
• pumping up takes longer than 160 seconds (
on the display)
• there is a system or device error (
appears on the display)
• the batteries are almost empty
In the above cases, you must repeat the measurement.
Make sure that the cuff tube is properly inserted and that
you do not move or talk.
Re-insert the batteries if necessary, or else replace them.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section "Technical speci-
fications", the technical alarm will appear on the display
indicating either " " or "Lo Lo". In such cases, you should seek
medical assistance and check the accuracy of your proce-
dure.
The limit values for the technical alarm are factory set and
cannot be adjusted or deactivated. These alarm limit val-
ues are accorded second priority under the standard IEC
60601-1-8.
or
appears
,
,
or
.
29
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear auto-
matically after about 8 seconds.

9. Cleaning and storing the device and cuff

• Clean the device and cuff carefully using a slightly damp
cloth only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under
water, as this can cause liquid to enter and damage the
device and cuff.
• If you store the device and cuff, do not place heavy ob-
jects on the device and cuff. Remove the batteries. The
cuff line should not be bent sharply.

10. Technical specifications

Model no.
BM 47
Measurement
Oscillometric, non-invasive blood
method
pressure measurement on the upper
arm
Measurement range
Cuff pressure 0-300 mmHg,
systolic 60-260 mmHg,
diastolic 40-199 mmHg,
Pulse 40-180 beats/minute