IMPORTANT! Do not provide these setup instructions to the patient. Only qualified medical professionals
should access these device settings.
Follow these instructions to access the device settings and prepare the device for patient use. All settings
have default values as shown below.
First -
With the device powered and
turned off, press and hold the
and
DELAY
buttons.
Second -
While holding the two buttons,
press the
ON/OFF
button. The display will
show "Clinical Menu". You can now
access the Clinical Setup Menu.
Clinical Menu Order
CPAP
Bilevel S
Bilevel ST
Bilevel T
Menu Item
Description
Mode
Mode
Mode
Mode
Language
Select device language. Range: English, French, German, Spanish, Italian, Dutch,
1
1
1
1
Selection
Danish, Finnish, Norwegian, Swedish.
Adjust operating mode as prescribed by doctor.
Operating
2
2
2
2
DV55 Range: CPAP , Bilevel S
Mode
DV56 Range: CPAP , Bilevel S, Bilevel ST, Bilevel T
Set
Set to CPAP pressure prescribed by doctor.
3
N/A
N/A
N/A
Pressure
Range: 3-25 cmH
(CPAP)
Set to IPAP pressure prescribed by doctor. Range: EPAP-25 cmH
N/A
3
3
3
IPAP
0.5. The pressure delta between IPAP and EPAP is also displayed for reference.
Set to EPAP pressure prescribed by doctor. Range: 3-IPAP cmH
N/A
4
4
4
EPAP
0.5. The pressure delta between IPAP and EPAP is also displayed for reference.
Select period of time to stay at delay pressure. This delay is patient activated.
4
5
5
5
Delay Time
Range: 0-45 minutes, increments of 5
If prescribed, set to a level that is comfortable for the patient to fall asleep. During
Delay
a delay the pressure starts at this level and rises to the set CPAP/IPAP/EPAP
5
6
6
6
Pressure
pressure over the delay time. Range: 3 cmH
increments of 0.5
IPAP
Pressure waveform profile (if in bilevel operating modes). Range: 0 - 5,
N/A
7
7
7
Rounding
increments of 1.
EPAP
Pressure waveform profile (if in bilevel operating modes). Range: 0 - 5, increments
N/A
8
8
8
Rounding
of 1.
Inspiratory
Adjust to control inhale trigger sensitivity. Range: 1 - 10, increments of 1. A setting
N/A
9
9
N/A
Trigger
of 1 is most sensitive and 10 least sensitive.
Sensitivity
Expiratory
Adjust to control exhale trigger (cycle) sensitivity. Range: 1 - 10, increments of 1.
N/A
10
10
N/A
Trigger
A setting of 1 is most sensitive and 10 least sensitive.
Sensitivity
This setting along with the I/E Ratio may control the devices backup rate. A device
Backup
N/A
N/A
11
9
generated breath will be triggered if the patient's spontaneous breath is slower than
Rate
this setting. Range: 4 – 30 bpm, increments of 1.
This is the ratio of the inhalation to exhalation time and is expressed as 1:X. This
N/A
N/A
12
10
I/E Ratio
setting may affect the devices backup rate. Range: 1:0.5 to 1:4 with increments of
0.1.
Pressure
The device operating pressure may be displayed in one of three pressure units:
6
11
13
11
Unit
cmH
O, hPa or mbar.
2
The device is supplied with air supply tubing of 1.8 m (6 ft) in length. The device
Tubing
7
12
14
12
may also be operated with air supply tubing of 3 m (10 ft). This setting should be
Length
updated to match the tubing length for maximum pressure accuracy.
Usage
The SmartCode feature tracks the number of days that the device was used greater
8
13
15
13
Threshold
than or equal to "usage threshold" hours. Range: 4 or 5 hours.
DeVilbiss Healthcare • 100 DeVilbiss Drive • Somerset, PA 15501 • USA • 800-338-1988 • 814-443-4881
Quick Set-Up Guide, DV55 and DV56 Series
Use the
LEFT
DOWN
button to navigate through the
menu.
Use the
UP
button to adjust the settings.
O
2
O, increments of
2
O, increments of
2
O to EPAP (or CPAP in CPAP mode),
2
(IntelliPAP
®
Clinical Menu Order
CPAP
Bilevel S
Bilevel ST
Mode
Mode
Mode
9
14
16
button and
RIGHT
10
15
17
11
16
18
button and
DOWN
Enable Submenu
Menu Item
Description
This option controls the Auto-OFF feature, which automatically stops the flow of air after taking off the mask. Range:
Auto-OFF
Enabled, Disabled
This option controls the Auto-ON feature, which automatically starts the flow of air after breathing once or twice into
Auto-ON
the mask. Range: Enabled, Disabled
Factory Default
Displays a reminder to the patient to check the fit of the mask if airflow was greater than 95L/Min for more than 10%
Mask Fit Chk
of the previous therapy session (including vent leak). The previous therapy session must be greater than 3 hours or a
English*
reminder will not be displayed. Range: Enabled, Disabled
DV55: Bilevel S
Notifications
Setting to enable mask/filter check notification. Range: Enabled, Disabled
DV56: Bilevel ST
Low (Standby)
This option allows the low (standby) keypad and display backlight to be enabled or disabled. Range: Enabled, Disabled
Backlight
15.0 cmH
O
Exit
2
Press the button to exit and return to Clinical Menu.
Submenu
15.0 cmH
O
2
Lockout Submenu
Menu Item
Description
8.0 cmH
O
2
Delay Time
Used to allow patient access to CPAP delay time setting. Range: Locked, Unlocked
Lockout
20 min
Tube Length
Used to allow patient access to tubing length setting. Range: Locked, Unlocked
Lockout
Rounding
Used to allow patient access to IPAP Rounding and EPAP Rounding settings (bilevel operating modes only). Range:
4.0 cmH
O
2
Lockout
Locked, Unlocked
Triggering
Used to allow patient access to Inspiratory Trigger Sensitivity and Expiratory Trigger Sensitivity settings (bilevel S and ST
Lockout
operating modes only). Range: Locked, Unlocked
3
Auto-ON/OFF
Used to allow patient access to Auto-ON and Auto-OFF settings. Range: Locked, Unlocked
Lockout
3
Exit Submenu
Press the button to exit and return to Clinical Menu.
5
* DV55NE and DV56NE series SleepCube devices have factory defaults of hPa for Pressure Unit, German for Language Selection, and Locked for Tube Length Lockout.
Other Settings
5
Filter Cleaning Reminder – Displays a reminder to the patient to clean the filter every 50 hours of use, or approximately every 10 days. This feature is
controlled by the notifications setting and is preset to enabled.
10 bpm
Other setup information
• This device features automatic altitude adjustment for altitudes up to 2600m (8500ft) and needs no manual adjustment.
• This device features a pressure sensor therefore no fine pressure adjustments are needed.
1:2.0
• The SmartLink Data Management System may be used for additional reminders and notifications.
• Clear compliance meter and SmartCode data by pressing button for eight seconds with compliance meter shown (value will go to zero).
cmH
O*
2
All device settings are immediately updated in the device as you make adjustments. When finished press the ON/OFF button once to exit the Clinical Setup
Menu.
1.8 m (6 ft)
4 hours
Sunrise Medical Ltd. • Sunrise Business Park • High Street • Wollaston, West Midlands • DY8 4PS • England • 44-138-444-6688
❙
Bilevel • SleepCube
Bilevel)
®
Bilevel T
Menu Item
Description
Mode
Cumulative hours of "while breathing" time. Clear compliance meter and
Compliance
SmartCode data by pressing button for eight seconds with compliance meter
14
Meter
shown (value will go to zero).
Enable
15
Press the button to enter. Reference the Enable Submenu section for details.
Menu
Lockout
16
Press the button to enter. Reference the Lockout Submenu section for details.
Menu
EN
Factory Default
N/A
N/A
N/A
Factory Default
Disabled
Enabled
Enabled
Enabled
Enabled
N/A
Factory Default
Unlocked
Unlocked*
Locked
Locked
Unlocked
N/A
©2009 DeVilbiss Healthcare 10.09