Communication Commands (continued)
Command Function
OFF
Turns balance OFF.
PSN
Show Serial Number.
PV
Version: Print product name, software revision and LFT ON (if LFT is set ON).
PU
Print current mode/unit
x#
Set PC ref wt (x) in grams
P#
Print PC ref wt
x%
Set % ref wt (x) in grams
P%
Print % ref wt

6.3.2 RS232 (DB9) Pin Connections

Pin 2: Balance transmit line (TxD)
Pin 3: Balance receive line (RxD)
Pin 5: Ground signal (GND)
Pin 7: Clear to send (hardware handshake) (CTS)
Pin 8: Request to send (hardware handshake) (RTS)
6.3 Compliance
Compliance to the following standards is indicated by the corresponding mark on the product.
Mark
Standard
This product conforms to the EMC Directive 2004/108/EC, the Low Voltage Directive
2006/95/EC and the Non-Automatic Weighing Directive 2009/23/EC. The complete
declaration of Conformity is available from Ohaus Corporation.
AS/NZS4251.1; AS/NZS4252.1
CAN/CSA-C22.2 No. 1010.1-92; UL Std. No. 3101-1
Important notice for verified weighing instruments
Weighing Instruments verified at the place of manufacture bear one of the preceding
marks on the packing label and the green 'M' (metrology) sticker on the descriptive
data plate. They may be put into service immediately.
Weighing Instruments to be verified in two stages have no green 'M' (metrology) on the
descriptive data plate and bear one of the preceding identification marks on the
packing label. The second stage of the initial verification must be carried out by an
authorized and certified service organization established within the European
Community or by the National Notified Body.
The first stage of the initial verification has been carried out at the manufacturer's work.
It comprises all tests according to the adopted European Standard EN 45501:1992,
paragraph 8.2.2.
If national regulations limit the validity period of the verification, the user of the
weighing instrument must strictly observe the re-verification period and inform the
respective Weights and Measures authority.