Technical Data - Dräger X-am 1 Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 6
Performing the measurement
Connect the Dräger sampling hoses or the Dräger probes
(see "Order List" on page 5.) to the hose connection of the
pump.
Perform the measurement according to the Instructions for
Use of the gas measuring device.
Ending operation
Press the release key.
— Both LEDs flash green shortly and then red, accompanied by
a short signal tone.
— The pump is switched off.
Remove gas measuring device from the pump.
Observe the following during measuring mode with
pump
The required waiting time when flushing the hose probe:
Before every measurement, flush the Dräger sampling hose
or the Dräger probes with the air sample to be measured.
— A flushing phase is necessary to eliminate or minimize all ef-
fects associated with the use of a sampling hose or a probe,
e.g. memory effect, dead volume.
— The duration of the flushing phase depends on factors such
as e.g. type and concentration of the gas or vapor to be mea-
sured, material, length, diameter, and age of the sampling
hose or probe. Generally, when using a sampling hose
(brand new, dry, clean), a typical flushing time of approx.
3 seconds is required for each meter. This flushing time ap-
plies in addition to the sensor response time (see the Instruc-
tions for Use of the used gas measuring device).
Example:
— In the case of a sampling hose with a length of 10 m, the flus-
hing time is approx. 30 seconds and the sensor response
time is in addition approx. 60 seconds. Therefore, the total
time before reading the gas measuring device is approx.
90 seconds.
— The flow rate alarm is delayed by 10 to 30 seconds depen-
ding on the length of the hose.
Displays, Alarms, Remedies
LED / Signal Tone
Alarms
Remedies
Bat LED flashes
Battery alarm
Change batteries or
red, accompanied
charge rechargeable
by a signal tone
batteries.
Flow LED flashes
Flow alarm
Check sampling hose
red, accompanied
or probe for blockages.
by a continuous
Change filter on intake
tone
port
LED / Signal Tone
Alarms
Remedies
Both LEDs flash
Device alarm
Have fault corrected by
red, accompanied
Dräger Service.
by a signal tone
Replacing the Batteries
Remove the gas measuring device from the device recepta-
cle before replacing the batteries, see "Ending operation".
Replace batteries,see "Commissioning" on page 4.
Changing the Filter
Unscrew the filter in a counterclockwise direction.
Unscrew the connecting bush from the filter.
Screw the connecting bush onto the new filter.
Screw the filter in a clockwise rotation onto the device.
Storing the Pump
Store the pump in a clean and dry place.
— It is recommended to store the pump separately from the gas
measuring device to ensure the flexible seal of the pump re-
tains its shape and sealing force.
In the case of long storage periods:
Remove batteries from the pump if the pump is not used for
more than 6 months. This avoids damage to the pump cau-
sed by leaking batteries.
Disposing of the Device
This product must not be disposed of as municipal
waste. This is indicated by the adjacent icon.
You can return this product to Dräger free of
charge. For information please contact the national
sales organisations and Dräger.

Technical Data

Ambient conditions
o
o
during operation
–20
C to 50
C
10 % r.h. to 90 % r.h.,
non-condensing
700 hPa to 1300 hPa
o
o
storage
–20
C to 50
C
10 % r.h. to 70 % r.h.,
non-condensing
700 hPa to 1300 hPa
Protection class
IP 54
Operation time
30 hours, with brand new alkaline batte-
ries, depending on the operating tem-
perature and storage conditions
Power supply
3 alkaline batteries – T4 – (not rechar-
geable!) Energizer No. E91, Energizer
No. EN91 (Industrial), Varta Type 4106
(power one), Varta Type 4006 or
3 NiMH rechargeable batteries – T3 –
(rechargeable) GP 180AAHC
(1800 mAh) max. 40
perature
Flow rate
approx. 0.5 L/min
Maximum hose length 30 m
Alarms
optically by LEDs, acoustic signal tone
Dimensions
62 mm x 82 mm x 240 mm (W x H x D)
Weight
approx. 400 g
Approvals
(see "Notes on Approval" on page 46)
Order List
Designation and Description
Dräger Pump X-am 1/2/5000
Accessories / Spare Parts
Alkaline batteries "AA" batteries
Connection set filter
Complete cap
Sealing cap
Measuring probe 0.5 m
Measuring probe 1.5m
Plug-in telescopic probe
Telescopic probe 100 with accessories
Telescopic probe 150 stainless steel
Tester 90
Float probe with accessories
Fluoroelastomer hose
Hose
Case pump X-am 1/2/5000
o
C ambient tem-
Part No.
83 19 400
83 18 708
83 19 364
83 19 356
83 19 363
64 08 238
64 08 239
68 01 954
83 16 530
83 16 533
83 16 532
83 18 371
12 03 150
11 80 681
83 19 385
5

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

X-am 2X-am 5000

Table des Matières