U .S . Federal law restricts this device to sale by or on the order of a physician . This device SHOULD
NOT be used without proper training and preceptorship .
For monopolar systems it is recommended to keep the voltage/power as low as possible to achieve the
desired end effect (due to the potential for capacitive coupling and inadvertent burning at high voltages) .
Adequate anesthesia is indicated for all electrosurgical procedures .
Inspect loop and ball for adequate insulation .
Avoid using an electric extension cord (danger of separation) .
Use a smoke evacuator with this device .
This device generates high frequency energy that could, under some circumstances, cause interference
to other equipment .
For situations where the MAXIMUM OUTPUT VOLTAGE is less than or equal to 1600V, ASSOCIATED
EQUIPMENT and ACTIVE ACCESSORIES should be selected that have a RATED ACCESSORY
VOLTAGE equal to or greater than the MAXIMUM OUTPUT VOLTAGE .
-
ASSOCIATED EQUIPMENT and ACTIVE accessories should be selected with RATED
ACCESSORY VOLTAGE ≥ MAXIMUM OUTPUT VOLTAGE when the smaller of variable y
(see below) or the number 6 is ≤ CREST FACTOR for that HF SURGICAL MODE.
-
When MAXIMUM OUTPUT VOLTAGE (Umax) is > 1600 V, and the CREST FACTOR is < the
variable y calculated below, indicating that any ASSOCIATED EQUIPMENT and ACTIVE
ACCESSORIES used in such mode or setting must be rated to withstand the combination of actual
voltage andm CREST FACTOR .
U
max
y =
600 [Volts]
No modification of this equipment is allowed .
No customer-serviceable parts .
Only replacment of power cords or fuses can be done by the user . Only replace with these items with
the exact replacement part(s) available from CooperSurgical .
Warning: To avoid risk of Electrical Shock, this equipment must only be connected to a supply mains with
protected earth .
Section 5 Professional Use Guide

5.1 Indications for OB/GYN Applications

Cervical conization
External anogenital lesion
Large vaginal intraepithelial neoplastic (VAIN) lesions
Large Loop Excision of the Transformation Zone (LLETZ) in the diagnosis and treatment of some
cervical intraepithelial neoplasias (CIN) and dysplasias
– 400 [Volts]
CAUTIONS
10