Error Codes (RH digital)
Error
Cause
code
E3
Temperature inside device is too high
E4
Motor or magnetic rods blockage
E11
Break in safety circuit
E21
Failure in safety circuit test
E22
E29
E24
Surface temperature (Temperature of contr- Heating off - Switch off device until the surface temperature of the heating plate is lower than
ol sensor) of the heating plate is higher than the
set safe temperature limit
E25
Heating and switching element monitoring
E26
Difference between temperature of safety Heating off - Switch off device.
sensor and temperature of control sensor
Control temperature > (Safety temperature + 40 K)
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Please contact the service department;
- Send the device for repair, including a short description of the fault.
Technical data
Device
Operating Voltage Range - Rated voltage
Frequency
Power consumption (+10%) max.
Permissible duration of operation
Permissible ambient temperature
Permissible relative humidity
Protection type acc. DIN EN 60529
Protection class
Overvoltage category
Contamination level
Operation at a terrestrial altitude
Dimensions (W x D x H)
Weight
Motor
Speed range
Power consumption
Setting resolution (digital/basic)
Speed stability (no load) at rated voltage
Stirred quantity max.(H
2
Heating plate
Dimensions (Ø)
Material
Heating
Heating power (-5%/+10%) at rated voltage
Adjustment and display resolution
Surface temperature
Hysteresis of heating plate at rated voltage
no container, center of heating plate at 100 °C
Electronic Thermometer
Adjustable safety circuit
Safety temperature limit (adjustable)
Effect
Heating off - Switch off device and allow to cool down.
Heating off - Switch off device.
Motor off
Heating off - Plug in contact plug (N).
Heating off - Switch off device.
O)
(digital/basic)
Solution
- Warning! Only to be carried out by authorized service personnel:
Carry out an internal test on the device to check the plug-in connector for the motor.
- Plug in contact thermometer/temperature sensor.
- Replace faulty connecting cable, plug, or contact thermometer.
- Switch off and switch on again after about 1 minute.
When fault is indicated again, please contact with IKA
- Set a higher safe temperature limit.
the selected safe temperature limit.
- Set a higher safe temperature limit.
- Safety temperature limit > 100 °C
See also "Functional check of inactivating the safety circuit".
- Warning! Only to be carried out by authorized service personnel:
Carry out an internal test on the device to check the plug-in connector for the heating element.
- Warning! Only to be carried out by authorized service personnel:
Carry out an internal test on the device to check the plug-in connector for the temperature sensor.
Vac
230 ± 10% / 115 ± 10% / 100 ± 10%
Hz
50 / 60
W
620
%
100
°C
+5 to +40
%
80
IP 21
I
II
2
m
max. 2000
mm
250 x 160 x 100
kg
2.8
rpm
0; 100 - 2000
W
20
rpm
50/Scale
%
± 10
ltr
15
mm
135
stainless steel
W
600
K
5/Scale
°C
ambient temperature ...320
K
± 20
DIN 12878
°C
100 - 360
service.
®
15