Sanitas SBM 38 Mode D'emploi page 52

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 21
Recommended separation distances between portable and mobile RF communications equipment and the EQuIPMEnT or sYsTEM
– for EQuIPMEnT or sYsTEM that are not lIFE-suPPORTInG
Recommended separation distances between portable and mobile RF communications equipment and the SBM 38 blood pressure
monitor
The SBM 38 blood pressure monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are con-
trolled. The customer or the user of the SBM 38 blood pressure monitor can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the SBM 38 blood pressure monitor
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be esti-
mated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
nOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
nOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
separation distance according to frequency of transmitter
150 kHz to 80 MHz
d =
0.117
0.370
1.170
3.700
11.7
(m)
80 MHz to 800 MHz
d =
0.117
0.370
1.170
3.700
11.7
52
800 MHz to 2.5 GHz
d =
0.234
0.740
2.340
7.400
23.4

Publicité

Table des Matières
loading

Table des Matières