Manufacturer's Declaration
The intraoral camera emits electromagnetic energy at levels compliant to international stan-
dards; however, interference between the camera and other devices is possible if located in
close proximity. The customer or the user of the camera can help prevent electromagnetic inter-
ference by locating sensitive devices away from the camera.
This camera, classified as Medical Electrical Equipment, requires special precautions regarding
EMC and must be installed and put into service according to the EMC information provided in
the accompanying product documentation. Portable and mobile RF communications equipment
can effect Medical Electrical Equipment. This camera complies with EMC requirements when
used with the cables and accessories supplied with the product. The use of accessories and
cables other than those sold by DEXIS, LLC and specified as replacement parts for internal com-
ponents, may result in increased emissions or decreased immunity of this camera. This camera
should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is
necessary, this camera should be observed to verify normal operation in the configuration in
which it will be used.
Guidance and manufacturer's declaration - electromagnetic emissions for all
EQUIPMENT and SYSTEMS
Guidance and manufacturer's declaration – electromagnetic emissions
This camera is intended for use in the electromagnetic environment specified
below. The customer or the user of this camera should assure that it is used in such
an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
Electromagnetic environment -
guidance
Group 1
This camera uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
Class A
This camera is suitable for use in all
establishments other than domestic
and those directly connected to the
Class A
public low-voltage power supply
network that supplies buildings used
for domestic purposes.
Complies