Télécharger Imprimer la page

Agilent Technologies TwisTorr 304 FS AG Notice De Mode D'emploi page 272

Masquer les pouces Voir aussi pour TwisTorr 304 FS AG:

Publicité

17
Technical Information
Programming
272/301
R1 (R2) TYPE: selects the type of reference measurement to
drive the relay. Possible types:
Frequency: driving frequency of the pump
Power: power supplied to the pump
Time: time from start
Normal: achievement of normal condition
Pressure: pressure read by the gauge, if present
R1 (R2) SET VALUE: Sets the reference value at which the
relay R1 (R2) is activated. This parameter has no effect if
NORMAL (previous parameter) is selected as SETPOINT
TYPE.
R1 (R2) HISTERESYS: Sets, in percentage, the hysteresis of
the relay
R1 (R2) MASK TIME: sets the masked time (in hh:mm:ss),
measured from start, within which the SETPOINT R1 (R2) is
not driven even if the threshold value has been reached. This
parameter is not available if SETPOINT TYPE is set to
NORMAL
R1 (R2) OUTPUT LOGIC: Selects the control logic of the
SETPOINT output.
positive logic (the relay is energized when the
measurement exceeds the reference)
negative logic (the relay is energized when the
measurement is below the reference)
TwisTorr 304 FS AG Rack Controller User Manual / 87-901-029-01 (E)

Publicité

loading

Ce manuel est également adapté pour:

X3506-64002X3506-64003