Where:
is the gauge output voltage;
V
g
V
out
V
g_min
values can be drawn from the gauge technical specifications or
from the following table.
ID
1
CT-100
2
FRG-700/702
3
IMG-500
4
PVG-500/502
6
PCG-750/752
9
CDG-500 T1000
10
CDG-500 T0100
11
CDG-500 T0010
12
CDG-500 T0001
TwisTorr 304 FS AG Rack Controller User Manual / 87-901-029-01 (E)
is the Analog Output voltage;
is the gauge voltage measuring range. These
V
÷
g_max
Gauge Model
Technical Information
Programmable Analog Output Diagrams
Pressure range
[mbar]
1.33E-2 to 1.01E3
5E-9 to 1E3
2E-9 to 1E-2
5E-4 to 1E3
5E-5 to 1.5E3
2.66E+1 to 1.33E3
2.66E+0 to 1.33E2
2.66E-1 to 1.33E1
2.66E-2 to 1.33
Vg-min ÷
Vg-max [V]
1.11 ÷ 8.25
1.82 ÷ 8.6
1.8 ÷ 8.5
1.9 ÷ 10
0.61 ÷ 10.23
0.2 ÷ 10
0.2 ÷ 10
0.2 ÷ 10
0.2 ÷ 10
225/301