P1 – Input
Figure 9
All the logic input to the controller must be connected at J1
mating connector. With the provided P1 mating connector
(shipped with pin 3 and pin 8 shorted) make the connections
with AWG 24, (0.24 mm2) or smaller wire to the pins indicated
in the figure to obtain the desired capability.
The following table describes the signals available on the
connector.
Tab. 3
Pins
Name
1-com
Start/Stop
2-com
Low Speed
3-com
Interlock
4-com
System
Override
TwisTorr 304 FS AG Rack Controller User Manual / 87-901-029-01 (E)
P1 input connector
Description
Contact closed = Start
Contact open = Stop
Contact closed = Low Speed
Contact open = High Speed
Contact closed = Active
Contact open = Inactive
If active allows the pump running, if
inactive stops the pump.
Contact closed
Contact open
If the contact is open, it allows the
pump running; if the contact is closed,
it stops the pump and all the others
connected devices (Fan, Vent).
Technical Information
Controller Connection
Electrical
Note
Optically isolated
Active only in
Vlow, max = 10v
Remote Mode
Imax = 5mA
operation
Imin (on) = 2mA
Optically isolated
Active only in
Vlow, max = 10v
Remote Mode
Imax = 5Ma
operation
Imin (on) = 2mA
Optically isolated
Active in all
Vlow, max = 10v
modes
Imax = 5mA
Imin (on) = 2mA
Optically isolated
Active in all
Vlow, max = 10v
modes
Imax = 5mA
Imin (on) = 2mA
17
215/301