Alarms

Alarm silence

This control switches off the mains power fault
and malfunction alarms.
Press this button to silence the alarms.
Audible alarm is re-activated automatically after
about 15 minutes.
The alarm can be stopped again for 15-minute
periods until the problem is solved. The indicator
light remains on. Refer to the Service Manual for more details.

Mains power fault

The visual alarm yellow
in the event of a power failure, or during transport. A low-priority audible alarm
sounds immediately, intermittently and at moderately strong volume.
The button
To stop the alarms, reconnect the control unit to the mains power supply, press
to silence the battery, or press

Malfunction

The yellow visual alarm
occurs. A low-priority audible alarm sounds immediately, intermittently and at
moderately strong volume.
170924(2) - EN - Primo
comes on if the unit is disconnected from the mains,
remains active.
comes on after about 15 minutes if a pressure fault
User Manual
Alarm silence
(see "Alarm silence" above).