Transport & storage
Attention:
Environmental temperature on storage:
-10 to +60°C.
Environmental temperature on storage if the
battery is connected to the charger:
+10 to +40°C*.
The battery needs to be fully charged at least
once in every 3 months, even if the wheelchair
is not used.**
EN26
* For long-term storage with battery:
10-30°C
** In extreme temperatures, the bat-
tery must be charged monthly.
Switch off push assist
3
Resetting
The Aura control performs continuous system
checks to verify the correct operation of the
wheelchair. In the event that a fault is found,
the push assist will be disabled for safety
reasons. The wheelchair's parking brake will
then automatically be activated.
In the event of faults the battery indicator will
start to flash. The number of flashing lamps
indicates the type of fault. See the table in
the 'Faults' chapter for the meaning. You can
resolve many of these faults by resetting the
wheelchair:
1. Switch the push assist off and wait 5
seconds.
2. Then switch on the push assist again.
If resetting does not help, or if the fault occurs
often, the wheelchair should be checked by
the dealer and may only be re-used if the
dealer considers that the system works well.

Faults

EN27