Capacitance (ACD-21SW / ACD-21SWC only)

Ranges
Resolution
4μF
1nF
40μF, 400μF
1nF
4mF
1nF
Minimum input range: >100nF
4nF, 40nF, 400nF ranges unspecified
When the capacitor to be tested is connected, if "dS.C" symbol indicates on
LCD, it means there is voltage existing in the tested capacitor and need to be
discharged before testing.
Overload protection: 600VDC or AC rms

Temperature (ACD-21SW / ACD-21SWC only)

Range
-35°C ~ 400°C, -30°F ~ 400°F,
Resolution
0.1°F, 0.1°C
±(1.0% + 2°F) 32°F ~ 400°F ±(1.0% + 1°C) 0°C ~ 400°C
Accuracy
±(2.0% + 6°F) -30°F ~ 32°F ±(2.0% + 3°C) -35°C ~ 0°C
Sensor type: K-type thermocouple
Overload protection: 30V Max

Continuity

Range: 400e
Resolution: 1e
Audible indication: Less than 25e
Response time: 500ms
Overload protection: 600VDC or AC rms

(NCV) Non-Contact Voltage Indicator

Detect voltage from 70V to 600VAC (50Hz ~ 60Hz)
Red LED and audible indicator
10
www.
.com
Accuracy
±(3.0% rdg + 15 dgts)
±(3.0% rdg + 5 dgts)
±(5.0% rdg + 20 dgts)