OHAUS Valor
1000 Series
4.3 CALIBRATION
Span Calibration:
Turn on the scale. When display flashes all segments ("888888") press and hold
ZERO and SELECT at the same time until "SET--- " is displayed at the end of power
on sequence. Press the SELECT key until "SET-F1" is shown on the display.
Press ZERO key to display "A____x" (x=divisions in grams:1,2,5,10.20.50).
Press ZERO key to accept the default division setting which is shown on the display.
Do not change default.
Display shows "b____x" (x=decimal point:0,1,2,3,4). Press ZERO key to accept the
default decimal point setting which is shown on the display. Do not change default.
Display shows "Cxxxxx" (xxxxx=Capacity of scale). Press ZERO key to accept the
default Capacity setting which is shown on the display. Do not change default.
Display shows "dx" (x=Zero tracking setting:0-9 from 0.125d to 1.25d). Press
ZERO key to accept the default setting which is shown on the display.
Press ZERO and display will show "Load 0". Make sure pan is empty then press
ZERO key to accept. Display will show "CAL 0" briefly, then "Load F" will be
displayed.
Press ZERO to display "Fxxxxx" (xxxxx=Span weight ). To accept default span
weight, put displayed weight on the scale, and press ZERO to accept.
If span weight is not available, user can select a lower span calibration point value.
Use the SELECT (scroll to next digit), TARE (decrease blinking digit) and M+
(increase blinking digit) keys to increment or decrement span calibration values.
Once desired span calibration weight is set, place displayed weight on the scale,
and press ZERO to accept. After completion, scale will be in weighing mode.
To abort calibration at any point, turn scale off. To restart calibration, follow
calibration procedure from beginning.