Table 2 Measurement warnings and errors (continued)
Error/Warning
Time > 60 s
Make sure that the membrane and the diaphragm are properly
immersed in the sample.
Time > 150 s
Examine the temperature.
Time > 300 s
Examine the probe: Clean the probe (refer to
on page 13 for more information); make sure that there are no
air bubbles in the membrane. Shake the probe like a
thermometer; connect a different probe to verify if problem is
with probe or meter.
10.389 mg/L
ISE measurement: The measured value is superior to the
>>>>>
highest standard used in calibration. Measure again.
00012 00:13
0.886 mg/L
ISE measurement: The measured value is inferior to the lowest
<<<<<<
standard used in calibration. Measure again.
00018 00:11

Replacement parts and accessories

Note: Product and Article numbers may vary for some selling regions. Contact the
appropriate distributor or refer to the company website for contact information.
Replacement parts
Description
sensION+ PH3 Lab pH-meter with accessories, without
probe
sensION+ PH31 Lab pH-meter, GLP, with accessories,
without probe
sensION+ MM340 Lab pH & Ion-meter, GLP, 2 channels,
with accessories, without probe
sensION+ EC7 Lab conductivity meter, with accessories,
without probe
16 English
Solution
Clean the probe
Item no.
LPV2000.97.0002
LPV2100.97.0002
LPV2200.97.0002
LPV3010.97.0002
Replacement parts and accessories (continued)
Description
sensION+ EC71 Lab conductivity meter, GLP, with
accessories, without probe
sensION+ MM374, 2 channel Lab meter, GLP, accessories,
without probes
Consumables
Description
pH buffer solution 4.01, with certificate, 125 mL
pH buffer solution 7.00, with certificate, 125 mL
pH buffer solution 10.00, with certificate, 125 mL
pH buffer solution 4.01, with certificate, 250 mL
pH buffer solution 7.00, with certificate, 250 mL
pH buffer solution 10.00, with certificate, 250 mL
pH buffer solution 4.01, with certificate, 1000 mL
pH buffer solution 7.00, with certificate, 1000 mL
pH buffer solution 10.00, with certificate, 1000 mL
Electrolytic solution (KCl 3M), 125 mL
Electrolytic solution (KCl 3M), 250 mL
Electrolytic solution (KCl 3M), 50 mL
Electrolytic solution 0.1 M, 125 mL
Enzyme solution
Pepsin Cleaning Solution
Electrode cleaning solution
Item no.
LPV3110.97.0002
LPV4110.97.0002
Item no.
LZW9460.99
LZW9461.97
LZW9470.99
LZW9463.99
LZW9464.97
LZW9471.99
LZW9466.99
LZW9467.97
LZW9472.99
LZW9510.99
LZW9500.99
LZW9509.99
LZW9901.99
2964349
2964349
2965249

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Sension+mm374Sension ph31

Table des Matières