Analog Connection To Transmitter Or Controller - Mettler Toledo InPro 6000 Série Instructions D'utilisation

Masquer les pouces Voir aussi pour InPro 6000 Série:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

InPro 6000 Series Optical O
VP8 cable InPro 6860 i / nA
Color
gray
blue
pink
brown RS485–
green/
yellow shield
For M400 2-w, M400 FF and M400 PA, the optical
oxygen sensor needs to be powered separately with
an appropriate power supply.

4.2.3 Analog connection to transmitter or controller

The InPro 6860 i is equipped with two different inter-
faces:
1. Digital communication to the transmitter and any
Modbus system.
2. Analog output (nA or 4 – 20 mA) plus a simulated
NTC 22 kΩ temperature signal. The mA version
includes HART communication.
a
Attention: Never connect an InPro 6860 i with mA
output to a nA input. This could damage the electron-
ics of the sensor and/or the instrument the sensor is
connected to.
VP6 cable for InPro 6860 i / nA or InPro 6860 i/ mA
Color
black / transp
red
gray
blue
white
green
green / yellow
Additionally to this manual the following documents
and tools are available:
– Modbus Programmers Guide
– Optical O
– mA-Output configuration tool
The materials can be downloaded here:
www.mt.com/pro-ODO-documentation
© 01/ 2016 Mettler-Toledo GmbH InPro 6000 optical O
Printed in Switzerland
2
M400
Function
TB4
24 VDC+
1
24 VDC–
2
RS485+
8
7
4
InPro 6860 i / nA
Function
cathode (nA)
anode (nA)
24 VDC, 100 mA+
24 VDC, 100 mA–
NTC 22 kΩ
NTC 22 kΩ (GND)
shield
HART communication manual
2
Sensors 12 mm
M800
1/2 channel
TB2
9
10
14
13
12
M800
M400
4 channel
2w/FF/PA
TB2 or
TB2
TB4
9
n.c
10
n.c.
14
O
13
N
12
M
InPro 6860 i / mA
Function
4–20 mA+ HART
4–20 mA– HART
24 VDC+
24 VDC–
NTC 22 kΩ
NTC 22 kΩ (GND)
shield
sensors
2
52 206 256
17