Configure the network
D A N G E R
Electrocution Hazard. Always remove power from the instrument
before making any electrical connections.
The Profibus network card provides an interface for RS485 connection.
Before use, the network card must be configured for the location in the
network. Use the switch settings on the top of the network card for
configuration (refer to the Installation section).
1.
Termination switch–Termination Off. Set the switch to this position
if this is not the last slave on the bus.
2.
Termination switch–Termination On ("T" position). Set the switch to
this position if this is the last or only slave device on the bus.
Operation
User navigation
Refer to the controller documentation for keypad description and
navigation information.
Setup the network
When the Profibus network card is installed, the controller requires the
correct configuration of the device and data order.
Note: Refer to the controller documentation for keypad description, basic
navigation information and controller setup.
1.
Select SC1000 SETUP from the menu.
2.
Select NETWORK MODULES.
3.
Select PROFIBUS-DP.
4.
The Profibus configuration screen is displayed.
Figure 3 Profibus configuration menu - Telegram function
1
ENTER button—Saves
configuration and returns to the
FIELDBUS menu
2
CANCEL button—Returns to the
FIELDBUS menu without saving
3
ADD button—Adds new device/
tag to the Telegram
4
DELETE button—Removes a
device/tag from the Telegram
5
UP/DOWN arrow—Moves
device/tag up and down
English
6