PLC Configuration
Read 4 Byte from the sc1000 telegram to PLC
Modul "2 Words from Slave"
Input
Read 2 Byte from the sc1000 telegram to PLC
Modul "1 Word from Slave"
Input
Modul "empty slot"
place holder
Write 4 Byte from PLC Output to the sc1000
Modul "2 Words to Slave"
telegram
Write 2 Byte from PLC Output to the sc1000
Modul "1 Word to Slave"
telegram
"Exchange 4 Byte PLC Output and Input with
Modul "2 Words to/from Slave"
the sc1000 telegram
Exchange 2 Byte PLC Output and Input with
Modul "1 Word to/from Slave"
the sc1000 telegram
Troubleshooting
W A R N I N G
Multiple hazards. Do not disassemble the instrument for maintenance or service.
If the internal components must be cleaned or repaired, contact the manufacturer.
Error and status indicators
Error and status words follow the same standard definition for all sc
probes and controllers.
Table 3
lists bit position and error messages.
and status messages.
A bit value of zero shows the error or status condition that is not true.
A bit value of 1 shows the error or status condition that is true. For
example, if Bit 0 has the value of 1, an error has occurred during the last
calibration.
English
13
Table 4
lists bit position
Table 3 Error messages
Bit
Message
0
Measurement calibration error
1
Electronic adjustment error
2
Cleaning error
3
Measuring module error
4
System re-initialization error
5
Hardware error
6
Internal communication error
7
Humidity error
8
Temperature error
9
10
Sample warning
Questionable calibration
11
warning
Questionable measurement
12
warning
13
Safety warning
14
Reagent warning
15
Maintenance required warning
Indication
An error has occurred during the last
calibration
An error has occurred during the last
electronic calibration
The last cleaning cycle failed
A failure has been detected in the
Measurement Module
Some settings are inconsistent and have
been reset to factory defaults
A general hardware error has been
detected
A communication failure within the
device has been detected
Excessive humidity has been detected
within the device
Temperature within the device exceeds
a specified limit
Some action is required with the sample
system
The last calibration may not be accurate
One or more of the device
measurements are out of range or are of
questionable accuracy
A condition has been detected which
may result in a safety hazard
The reagent system requires attention
The device requires maintenance