Hach 2100Q Manuel D'utilisation De Base page 15

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 31
Error/Warning
Description
Please check the lamp!
Signals are too low on
the 90° and 180°
detector.
Temperature too high!
Temperature has
Switch off instrument.
exceeded the meter
limits (>60 °C or
>140 °F).
RST: Average value!
Solids are settling too
slowly. The reading
mode is not suitable for
this sample.
Confidence level is <
The reading mode
95%
Rapidly Settling Turbidity
did not meet the range of
≥ 95% confidence.
Standard value out of
Used incorrect standard
range. Insert standard
value for the reading.
and push Read
ID already in use. Enter
The Operator or Sample
new ID
ID is unavailable as it is
already assigned.
Error - Security Please
No password is created.
set password before
activating security
Please enter at least
Password must contain
one character.
minimum of one
character.
Solution
2100Q: The lamp is
Password incorrect.
defective. Change the
Please retry.
lamp (refer to
Replace
Please disconnect the
the lamp
on page 12).
USB cable from your
2100Qis: Contact
computer.
technical support.
Turn off the meter and let
USB module memory
it cool down.
full. Delete data and try
again.
Select Normal or Signal
Average reading mode.
Delete Last Reading
Failed!
• Invert the sample
Delete Data Log failed!
several times so that
the solids allocate.
Can't read data set!
Repeat the reading
Can't store data!
again.
• Switch to the Normal
Can't store to the
reading mode if the
Reading Log!
sample is stable and
does not have settable
Can't store to the Verify
solids.
Cal Log!
Error storing data!
Insert the appropriate
standard and read again.
Error reading data!
Create a new ID.
Create a new password.
Create a password of at
least one character.
Error/Warning
Description
Incorrect password was
entered.
Data storage does not
respond while connected
to the meter and the
computer.
Data storage is full.
Error in the data storage.
Solution
Enter the appropriate
password.
Disconnect the USB
cable from the meter and
try sending data again.
1. Connect USB/power
module to the computer.
2. Download the stored
data to the computer.
3. Delete Data Log on
the module.
Turn the meter off and
on. If the error message
still occurs, contact
technical support.
English 15

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

2100qis

Table des Matières