AMI Soditrace
Dépannage
7.6.
A-96.250.552 / 080720
Remplacement des fusibles
AVERTISSEMENT
External Voltage.
External supplied devices connected to relay 1 or 2 or to the
alarm relay can cause electrical shocks.
Make sure that the devices connected to the following con-
tacts are disconnected from the power before resuming in-
stallation.
– relay 1
– relay 2
– alarm relay
When a fuse has blown, find out the cause and fix it before replac-
ing it with a new one.
Use tweezers or needle-nosed pliers to remove the defective fuse.
Use original fuses provided by SWAN only.
A
B C D E F
A
Version AC: 1.6 AT/250 V Alimentation électrique de
l'instrument
Version DC: 3.15 AT/250 V Alimentation électrique de
l'instrument
B
1.0 AT/250 V Relais 1
C
1.0 AT/250 V Relais 2
D
1.0 AT/250 V Relais d'alarme
E
1.0 AF/125 V Sortie 2
F
1.0 AF/125 V Sortie 1
G
1.0 AF/125 V Sortie 3