9 Maintaining your STATIM
Cassette
Biological Filter
Solenoid Valve
Pump
Check Valve
Water Reservoir
Calibration
Technician
Check the tray, lid and seal for damage. Replace if necessary.
Inspect the biological filter for moisture.
Inspect the valve and clean if dirty. Replace the plunger if defective.
Clean the filters, replace if dirty.
Remove the exhaust tube from the back of the unit during the air drying
phase. Check for air coming from the fitting.
Remove the air compressor tube from the check valve inlet while running
a cycle. Make sure no steam is leaking from the valve. Replace if there are
any leaks.
Check the reservoir for dirt. Clean and rinse with steam distilled water or
SciCan approved water sources only.
Calibrate the unit.