6 Storing and Retrieving Cycle Information
6.3 Cycle Printout Overview
Model: STATIM 2000 software: S203R604
Unit Identifier: autoclave has been set up as number 000
Water Quality: conductivity measurements
Cycle Counter: the number of cycles having been run on the unit = 2
Time / Date: 10:47 am 25th October 2007
Cycle Name: UNWRAPPED
Cycle Name cont'd - parameters: 135°C / 3.5 min.
Cycle Clock: starting at 0:00
Warm up complete: start of the conditioning phase is 1:19 (see cycle
graph - 'A' phase complete, start of 'B' phase)
Start Time of Pressurization: 1:27 (start of 'C' phase)
Start Time of Sterilization: 2:27 (start of 'D' phase)
Temp. / Press. & Time at start of sterilization ('D' phase)
Temp. / Press. & Time printed at 30 second intervals during
sterilization. ("D" phase)
Temp. / Press. & Time of end of sterilization phase (end of 'D' phase)
Time Venting started: 5:57 (start of 'E' phase)
Time Air Drying started: 6:42 (start of 'F' phase)
Cycle completion time: 22:42
Acceptable Tolerances
Sterilization time
Saturated Steam Pressure
Sterilization Temperature
STATIM 2000
UNIT #:
WATER QUALITY
CYCLE NUMBER
10:47
UNWRAPPED
135°C FOR 3.5 MIN.
CYCLE START
CONDITIONING
PRESSURIZING
STERILIZING
136.4ºC 220KPA
136.0ºC 219KPA
135.9ºC 222KPA
136.1ºC 222KPA
136.5ºC 225KPA
136.4ºC 225KPA
136.1ºC 221KPA
136.1ºC 221KPA
VENTING
AIR DRYING
CYCLE COMPLETE
"Sterilization time" (e.g. 3.5 mins) -0/+1%
202kPa - 240kPa for Unwrapped/Wrapped cycle
(103kPa - 131kPa for Rubber and Plastics cycle)
"Specified temp" -0/+4 (134°C-138°C) (121°C -125°C
for Rubber and Plastics cycle)
*data on Cycle Printout should fall within these ranges
38
S203R604
000
0.5 µS /0.3 ppm
000002
10/25/2007
0:00
1:19
1:27
2:27
2:27
2:57
3:27
3:57
4:27
4:57
5:27
5:57
5:57
6:42
22:42