PROCEDURE AP2.012
MOTOR SEAL REPLACEMENT (REFRIGERATED)
The motor seal is located in the bottom of the
bowl of the machine (fig 1).
Before starting, open the lid, switch off the
centrifuge and unplug power cable
CAUTION: THE FOLLOWING
OPERATIONS MUST BE PERFORMED
BY A QUALIFIED TECHNICIAN
Fig 1
Step 1: Remove the motor seal and
clean all around the bowl and motor throats
(silicone remaining)
Seal removed