Handling during measurement
Important:
The measuring axis
must be in central
position to the bore!
Important:
The measuring axis must
be in vertical position to
the bore!

Display adjustment

The gage should be verified before each series of measurements by
using a setting ring.
If there is a deviation from the specified measurement the device
should be adjusted by using the rotating scale (E).
The measurements have to be repeated until the gage is
correctly adjusted.
In order to increase the accuracy of the measurement, the gage should be
adjusted in the same position in which later the measurement will be effected.
8
Determination of the minimum measuring value via pivoting
the gage.
Measured value
too high
Determination of the maximum measuring value for measurement of
narrow grooves via lateral displacement of the gage, where the gage
cannot be pivoted vertically.
Measured value
too small
Measured value
correct
Measured value
correct
Measured value
too high
Measured value
too small
D
E