5. Use the following chart as a guideline to
determine the length of time the Sprint/Stroller/
Hi Flow Stroller will operate:
FCV Setting
Model
Sprint
Nominal
FCV
Sprint
Pos.
Off
30:00
0.25
18:00
0.5
12:12
0.75
9:18
1
7:30
1.5
5:30
2
4:18
2.5
3:36
3
3:06
4
2:24
5
2:00
6
1:42
8
N/A
10
N/A
12
N/A
15
N/A
Note: Times are in hours and minutes
(format 00:00).
Note: The "Nominal" times are for ideal conditions,
i.e. maximum fill, exact flow rates, good loss
rate, unit not being moved, etc.
These times are the maximum expected.
Note: Your individual results will vary.
6a.To verify the level of liquid oxygen in the unit
with the electronic liquid level gauge:
•
Gen 3 Gauge: Depress the push button (liquid
level switch) on top of the unit for two seconds
minimum. Read across the top of the light bar to
indicate contents level.
Sprint / Stroller / Hi Flow Stroller
Stroller
Hi Flow
Stroller
Nominal
Stroller
Hi Flow
Stroller
60:00
71:00
30:00
NA
23:00
32:00
17:24
NA
14:06
16:00
10:12
NA
8:00
8:00
6:36
6:20
5:36
5:20
4:24
4:00
3:36
NA
3:00
2:40
2:18
2:00
1:54
1:40
1:36
1:20
1:24
1:00
Gen 4 Gauge: Press the green operate button
•
on the face of the gauge. Read the arc of LEDs,
which indicates content level.
Caution: The Sprint/Stroller/Hi Flow
Stroller is empty if only the last segment
of the light bar is lit.
If the Low Battery indicator lights up when
the button is depressed, notify your health care
provider.
NOTE: The Sprint/Stroller/Hi Flow Stroller will
continue to supply oxygen even if the Low Battery
Indicator lights up, as long as there is liquid oxygen
in the unit.
6b. To verify the level of liquid oxygen in unit with
the mechanical liquid level gauge:
•
Support the unit, by the fabric strap, directly
above the black contents indicator spring scale
chamber.
•
Allow the unit to stabilize (i.e. not bounce up and
down), then read the exposed colored scale for
approximate liquid oxygen contents.
PN MN236-C4 C | User Manual
9 - ENG