Warning Information

Important: Read this manual thoroughly before
operating the Sprint/Stroller/Hi Flow Stroller.
RX Only.
WARNING: THIS DEVICE IS NOT IN-
TENDED FOR LIFE SUSTAINING USE.
WARNING: IF YOU FEEL THE EQUIPMENT IS NOT
OPERATING PROPERLY, CALL YOUR HEALTH CARE
PROVIDER. DO NOT ATTEMPT TO REPAIR OR
ADJUST THE UNIT YOURSELF.
WARNING: DO NOT MODIFY THIS EQUIPMENT
WITHOUT AUTHORIZATION FROM THE MANUFAC-
TURER.
WARNING: IF CONTINUITY OF OXYGEN SUPPLY IS
REQUIRED, ENSURE THAT AN ADEQUATE SUPPLY
OF OXYGEN AND/OR A SECONDARY OXYGEN
SUPPLY IS AVAILABLE AT ALL TIMES DURING
THERAPY.
WARNING: DO NOT ALLOW SMOKING, CANDLES,
OR OPEN FLAMES WITHIN 10 FEET (3 METERS) OF
THE DEVICE, OR CLOSER THAN 8 INCHES (20 CM)
FROM ANY SOURCE OF IGNITION.
WARNING: KEEP YOUR UNIT IN A WELL-VENTI-
LATED AREA.
WARNING: DO NOT STORE LIQUID OXYGEN
EQUIPMENT IN A CLOSET, CAR TRUNK, OR OTHER
CONFINED AREA. DO NOT PLACE BLANKETS,
DRAPERIES, OR OTHER FABRICS OVER EQUIP-
MENT.
WARNING: THIS PRODUCT CAN EXPOSE YOU
TO CHEMICALS INCLUDING NICKEL, WHICH IS
KNOWN TO THE STATE OF CALIFORNIA TO CAUSE
CANCER. FOR MORE INFORMATION, GO TO WWW.
P65WARNINGS.CA.GOV.
WARNING: IN THE EVENT THERE IS A SERIOUS
INCIDENT OCCURRING WITH THIS DEVICE, THE
USER SHOULD IMMEDIATELY REPORT THE INCI-
DENT TO THE PROVIDER AND/OR THE MANUFAC-
TURER. A SERIOUS INCIDENT IS DEFINED AS AN
INJURY, DEATH, OR POTENTIAL TO CAUSE INJURY/
DEATH SHOULD THERE BE A REOCCURRENCE OF
THE INCIDENT. THE USER CAN ALSO REPORT THE
INCIDENT TO THE COMPETENT AUTHORITY IN THE
COUNTRY WHERE THE INCIDENT OCCURRED.
Caution: Use the Sprint/Stroller/Hi Flow
Stroller only as directed by your doctor.
Caution: Federal law restricts this device to sale
by or on the order of a physician.
4 - ENG
PN MN236-C4 C | User Manual
Caution: The unit contains liquid oxygen which is
extremely cold, almost -300°F (-184°C). Exposure
to such a low temperature can cause severe
frostbite.
Caution: Liquid and gaseous oxygen, though
nonflammable, cause other materials to burn
faster than normal. This hazard, along with the
low temperature of liquid oxygen, warrants
certain safety precautions.
Caution: Keep flammable materials away from
this equipment. Aerosol sprays, oils and grease,
including facial creams and petroleum jelly, ig-
nite easily and may burn rapidly in the presence
of oxygen.
Caution: Smoking while wearing an oxygen can-
nula can cause facial burns and possibly result
in death.
Removing the cannula and placing it on clothing,
bedding, sofas, or other cushion material will
cause a flash fire when exposed to a cigarette,
heat source, spark or flame.
If you smoke please: (1) turn off the portable,
(2) take off the cannula, and (3) leave the room
where the device is located.
Caution: In the event of an accidental tip-over,
immediately but cautiously return the unit into
an upright position if possible. If any liquid oxy-
gen is escaping, leave the area immediately and
call your healthcare provider. Do not attempt
to move the unit or stop the liquid oxygen from
escaping.
NOTE: Do not touch frosted parts of any unit.
NOTE: Do not store or operate the portable coupled
to the stationary unit.
NOTE: Do not allow untrained personnel to handle or
operate this device.
NOTE: Use of this device is prohibited on commercial
passenger and cargo air flights by the Federal Avia-
tion Administration.
Intended Use
The CAIRE Sprint/Stroller/Hi Flow Stroller oxygen
unit is intended for the administration of supple-
mental oxygen. The device is not intended for life
support nor does it provide any patient monitoring
capabilities.