For the density determination of solids, use the low glass beaker (4) and one of the two holders for solids
which sink (6) or float (7).
Prepare the balance for the density determination (install draft
shield element, bracket, equalizing washer and platform) as
described in Section 2.
Suspend the supplied thermometer (8) from the edge of the low
beaker (4). Place the beaker on the pan and fill it with auxiliary
liquid (liquid of known density ρ
ethanol). Add enough liquid to ensure that the solid is covered by
at least 1 cm liquid after immersion.
Suspend the suitable holder (6) or (7) for solids which sink or float
from the bracket (the adjacent illustration shows the holder for
solids which sink).
Ensure that no air bubbles adhere to the immersed part of the holder
(remove any air bubbles by moving the holder or by means of a fine
brush).
Close the draft shield doors and tare the balance.
Place the solid in one of the two weighing pans of the bracket (2).
Wait until the weight display of the balance is stable (stability
detector fades) and note the displayed weight A (weight of sample
in air).
Note:
When solids with a weight greater than 20 g are weighed in air,
place in the pan at the top of the holder (above 20 g cornerload
errors may appear on one arm in the weighing).
Density determination kit for XS balances