The specified adjustment weight is displayed flashing.
Place the adjustment weight centrically on the weighing
pan and confirm with
signalling tone will sound. Adjustment will be performed.
Then the balance returns automatically into the weighing
mode.
An error message will be displayed in the event of an adjustment error or
incorrect adjustment weight. Remove the adjustment weight and repeat the
adjustment process.
Keep the adjustment weight close to the balance. Daily control of the
weighing exactness is recommended for quality-relevant applications.
EOB / EOE / EOS -BA-e-1741
. "CAL Y" will flash briefly and a
(example)
19