C16i
CAUTION
There is no warranty in case of un-lubricated compres­
sion pads and if the flexible tube has not been mounted
correctly!
4.2
Procedure for special mounting
situations
If there are extremely long distances between the sup­
ports of the load cells or if the weighbridges bend easily,
the rolling movements of the load cell produced by a load
can cause discrepancies in the results. This rolling mo­
tion is assisted by a lateral drift of the upper load cell load
introduction points under load, if the contact zone be­
tween the load introduction part and the load cell is far
below the neutral bending axis of the weighbridge. To
minimize the discrepancies which occur in these cases,
the load cells can be mounted at a slight slant, set to a
maximum of 1° inwards.