Service & maintenance information
Service and maintenance of a medical device is wholly the responsibility of the owner of that device.
Failure to service and maintain a device in accordance with the instructions may invalidate the
device's warranty. Furthermore, failure to service and maintain a device may compromise the clinical
condition or safety of users and/or their caregivers. Do not perform service and maintenance while the
user is in the product. Contact the local dealer for assistance, if needed, in setting up, using or how to
perform service and maintenance of the device.
Lifetime
The lifetime of this product in normal use
is 8 years if all maintenance and servicing
is carried out in accordance with the
manufacturer's instructions and demonstrably
recorded.
Reuse
This product and its accessories/components
are suitable for reuse.
Refurbishment for reuse
Before reuse or changing the user, follow the
service checklist, maintenance and washing
instructions and carry out regular inspections.
Service interval
The frequency of inspection may be
altered depending upon use and wear. It is
recommended that the product is inspected
yearly, every time the product is re-issued for
use and after a long-term storage (more than
4 months). The inspection should be carried
out by a person who understands the use of
the product.
Service checklist:
Check and adjust the following as per their
intended use:
• Cushioning
• Products equipped with a hydraulic pump
for height adjustment. Frame height can
be adjusted up and down
• Powered versions are able to retain power
after charging the battery
• For powered versions, reset the power unit
system using the handset by pressing the
up(f) and down(g) buttons simultaneously
for 5 seconds
• Products equipped with power for height
adjustment. The cables are in good
condition and the handset funtionalities
for height adjustment works as intended.
• Nuts and bolts on the product are
tightened
• Oil the swing-away parts
• Fittings and buckles for signs of worn out
parts
• Product for cracks or signs for worn-out
parts
• All labels on the product are intact
Repair or replace any damaged or worn out
parts.
EN
XX
Maintenance
Before every use it is recommended to:
Wipe off any residue and dirt from the product,
using a cloth with warm water and a mild
detergent/soap without chlorine and let it dry
before use carry out a visual inspection, for
any damaged or worn-out parts.
Washing
Main product
This product can be washed at 60° using a
mild detergent for 10 minutes in a washing
machine designed for medical devices. Use
the machine's drying function to dry the
product. Unmount all accessories and wash
them separately.
This product can be washed at 85° using a
mild detergent for 3 minutes in a washing
machine designed for medical devices. Use
the machine's drying function to dry the
product. Unmount all accessories and wash
them separately.
Before washing
Avoid degreasing of the piston rods, the
actuators should be retracted to minimum
stroke and without load before washing.
• Remove all accessories, unplug all cables
and make sure to dismount the battery,
control-box and handset. Leave the
actuators with the cables mounted on the
product (see page 22)
• Do NOT clean the power unit with a high
pressure cleaner or steam cleaner.
After washing
• Ensure correct placement of cables
inserted into the control box when
reinstalling the power unit system (see
page 23)
• Check that the handset functionalities
works correct
Handwash
This product and its accessories can also be
washed by hand. Use warm water and a mild
detergent/ soap without chlorine and let all
parts dry before use.
Disinfection
The product can be disinfected with a 70%
disinfectant IPA solution. It is recommended to
wipe off any residue and dirt from the product,
using a cloth with warm water and a mild
detergent/soap without chlorine and let it dry
before disinfection.
43