General Information; Instructions For Use; Preparation For Installation - Varian Turbo-V 1000 ICE Notice De Mode D'emploi

Masquer les pouces Voir aussi pour Turbo-V 1000 ICE:
Table des Matières

Publicité

GENERAL INFORMATION

This equipment is intended for use by professionals.
The user should read this instruction manual and any
other additional information supplied by Varian before
operating the equipment. Varian will not be held
responsible for any events occurring due to non-
compliance, even partial, with these instructions,
improper use by untrained people, non-authorized
interference with the equipment or any action contrary
to that provided for by specific national standards. The
Turbo-V
1000
ICE
microprocessor-controlled,
converters with self-diagnostic and self-protection
features.
The
controllers
drive
Turbo-V 1000 ICE pump during the starting phase by
controlling the voltage and current respect to the speed
reached by the pump. They incorporate all the facilities
required for the automatic operation of the Turbo-V
1000 ICE pump series.
Remote start/stop commands, pump status signals,
forepump
start/stop
commands,
signals (for pressure switch, water flow switch), heater
jacket control signals, flow meter control signals and
purge valve management signals are provided via
auxiliary connectors. The following paragraphs contain
all the information necessary to guarantee the safety of
the operator when using the equipment. Detailed
information is supplied in the appendix "Technical
Information".
This manual uses the following conventions:
!
The warning messages are for attracting the attention
of the operator to a particular procedure or practice
which, if not followed correctly, could lead to serious
injury.
The
caution
messages
procedures which, if not followed, could cause damage
to the equipment.
The notes contain important information taken from the
text.
series
controllers
solid-state,
frequency
(within
ten
steps)
interlock
control
WARNING!
CAUTION!
are
displayed
before
NOTE
STORAGE
When transporting and storing the controllers, the
following
environmental
satisfied:
temperature: from -20 °C to + 70 °C
relative humidity: 0 - 95% (without condensation)

PREPARATION FOR INSTALLATION

The controller is supplied in a special protective
are
packing. If this shows signs of damage which may have
occurred during transport, contact your local sales
office. When unpacking the controller ensure that it is
not dropped or subjected to any form of impact. Do not
the
dispose of the packing materials in an unauthorized
manner. The material is 100% recyclable and complies
with EEC Directive 85/399.
Each controller is factory set for a specific power
supply:
model 969-9447 is factory set for 220 Vac
operation
model 969-9547 is factory set for 120 Vac
operation
Check voltage selector window for correct set and
connect power cord.
65

INSTRUCTIONS FOR USE

requirements
Controllers packing
87-900-905-01 (I)
should