This service information describes the connection of the IntelliBridge to the Dräger Evita XL
ventilator.
Data communication
The IntelliBridge EC10 module supports RS232 serial data communication with the following
specifications for connection with the external device:
• Baud rate:
• Word length:
N O T E
RS232 parameters must be set and verified according to the Dräger Evita XL ventilator service manual
by authorized personnel.
Instruction for connection
Connect the IntelliBridge EC5 D-Sub 9-pin male connector to the RS232 connector (COM1) of the
Dräger Evita XL ventilator as shown in the photo below. Make sure to tighten the screws of the
connector properly.
IntelliBridge EC5 connection to RS232 ports
1
2
Service Information
7
19200
8 bits
1
Dräger Evita XL ventilator - RS232 (COM) port
IntelliBridge EC5 ID module
453564160841
• Stop bits:
• Parity:
2
7
1
Even
29