2. GENERAL DESCRIPTION
The instrument carries out the following measurements:
• DC and AC TRMS voltage up to 300V
• DC current from 0.1mA up to 10A
• AC TRMS current from 0.1mA up to 20A
• AC and DC leakage current
• Resistance and continuity test with buzzer
Each of these functions may be selected through a 6-position rotary switch, including the
OFF position and a key for enabling the HOLD function. The instrument also has a
"
"
MAX/MIN,
ZERO
appears on the LCD display with the indication of the measuring unit and of the enabled
functions.
2.1. MEASURING AVERAGE VALUES AND TRMS VALUES
Measuring instruments of alternating quantities are divided into two big families:
• AVERAGE-VALUE meters: instruments measuring the value of the sole wave at
fundamental frequency (50 or 60 Hz).
• TRMS (True Root Mean Square) VALUE meters: instruments measuring the TRMS
value of the quantity being tested.
With a perfectly sinusoidal wave, the two families of instruments provide identical results.
With distorted waves, instead, the readings shall differ. Average-value meters provide the
RMS value of the sole fundamental wave; TRSM meters, instead, provide the RMS value
of the whole wave, including harmonics (within the instruments bandwidth). Therefore, by
measuring the same quantity with instruments from both families, the values obtained are
identical only if the wave is perfectly sinusoidal. In case it is distorted, TRMS meters shall
provide higher values than the values read by average-value meters.
2.2. DEFINITION OF TRUE ROOT MEAN SQUARE VALUE AND CREST FACTOR
The root mean square value of current is defined as follows:
an alternating current with a root mean square value of 1A intensity, circulating on a
resistor, dissipates the same energy that, during the same time, would be dissipated by a
direct current with an intensity of 1A". This definition results in the numeric expression:
G=
The root mean square value is indicated with the acronym RMS.
The Crest Factor is defined as the relationship between the Peak Value of a signal and its
RMS value: CF (G)=
sinusoidal wave it is
as wave distortion increases.
"
"
"
and
key. For their use, please refer to § 4.2. The selected quantity
This value changes with the signal waveform, for a purely
2 =1.41. In case of distortion, the Crest Factor takes higher values